RL-9000 Patient Monitor

RL-9000 Patient Monitor

RL-9000 Patient Monitor


Technical Parameter:

ECG


* Lead selection: full lead, Ⅰ, , , aVF, aVR, aVL, V1-6 chest lead display

* Gain selection: ×1, ×2, ×0.25, ×0.5 four gears

* Frequency: 0.5Hz—100Hz

* Common mode signal rejection: greater than 70dB

* Scanning speed: 12.5mm/s, 25mm/s, 50mm/s three standard files

* Heart rate accuracy: ±5 beats/minute

* Calibration signal: 1mV (peak value, accuracy ±3%)

* ST segment detection: measuring range (-2.0mV---+2.0mV)

* Use standard: in line with ANSI/AMI EC13-2002 standard

* Response time: 1s

* Alarm mode: sound and light alarm

* Heart rate alarm preset range: (exceed the preset value alarm)

The upper limit is adjustable between 1-255 times/minute

The lower limit is adjustable between 0-254 times/minute


NIBP Blood Pressure


* Technology: oscillometric method

* Measurement method: reverse measurement data is more accurate

* Unit of measurement: mmHg/Kpa

* Overvoltage protection: double safety protection

* One measurement time: less than 45s

* Pressure range in cuff: 0-250mmHg

* Blood pressure measurement accuracy: ANSI/AAMI SP10-2002 standard stipulates the average error and standard deviation value

* Initial inflation of the cuff: 180mmHg (120mmHg for children)

* Cuff subsequent inflation: original systolic blood pressure +30mmHg

* Blood pressure measurement range and error:

Systolic blood pressure 60 mmHg~250 mmHg

Average pressure 45 mmHg~235 mmHg

Diastolic blood pressure 40 mmHg~220 mmHg

* Blood pressure measurement error: ±1.1KPa (±8mmHg) or ±10%, whichever is greater

* Response time: 1s

* Alarm mode: sound and light alarm

* Alarm prediction range and error: (Alarm when exceeding the preset value)

* Systolic blood pressure upper limit 1mmHg~255 mmHg

                        Lower limit 0 mmHg~254 mmHg

* Diastolic blood pressure upper limit 1mmHg~255 mmHg

         Lower limit 0 mmHg~254 mmHg

* Alarm error: ±1 mmHg

* Use standard: in line with ANSI/AAMI SP10-2002 standard


SPO2 Blood Oxygen Saturation


* Range: 0-100%

* Accuracy: 1%

* Resolution: 1%

* Pulse frequency: 20-250 beats/min

* Response time: 1s

* Alarm mode: sound and light alarm

* Alarm range and error: (Alarm when exceeding the preset value)

Upper limit 1%~100%

Lower limit 0%~99%


TEMP Body Temperature


* Technology: Highly sensitive thermistor probe

* Body temperature measurement range: 20°C -- 45°C

* Resolution: 0.1%

* Body temperature accuracy: ±1°C

* Response time: 1s

* Alarm mode: sound and light alarm

* Alarm preset range and error: (Alarm when exceeding the preset value)

The upper limit is adjustable between 20.1°C -- 45.5°C

The lower limit is adjustable between 20°C and 45.4°C

* Alarm error: ±1°C


RESP Breath


* Respiratory frequency range: 0-60 times/minute

* Accuracy: ±1rpm

* Response time: 1S

* Gain selection: ×1, ×2, ×4, ×0.5 four gears

* Alarm mode: sound and light alarm

* Alarm preset range and error: (Alarm when exceeding the preset value)

The upper limit is adjustable between 10-100 times/minute

The upper limit is adjustable between 0-99 times/minute

* Alarm error: ±1rpm


System


* Display: 10.1-inch TFT true color LCD display

* Battery (lithium battery): fully sealed maintenance-free battery

* Safety standard: IEC601-1 (International GB9706.1)

Inquiry:

CATEGORIES

LATEST NEWS

CONTACT US

Contact: Mr. Jin

Phone: + 86 15951972343

Tel: + 86 15951972343

Email: jin@runlai-medical.com; rl_medical@126.com

Add: ZhangJiaGang City, Jiangsu Province, P.R.China   PC:215600

Leave a message

Facebook

Twitter

LinkedIn

pinterest

Youtube

instagram

+ 86 15951972343

whatsapp

jin@runlai-medical.com; rl_medical@126.com

448473895

Top